Содержание
- 2. Reagents-Instructions for use Note: Check the inside of the neck of the wedge for foam after
- 3. Calcium The calibration interval (14 days maximum) is programmed on the RFID chip on the reagent
- 4. CO2 & ALP Keep the reagent tightly closed when not in use. When used in this
- 5. Total Protein The reagent is stable on-board in the refrigerated reagent area of the Medica EasyRA
- 6. The “Debug Mode” When Medica Technical Support says “Yes” or when additional troubleshooting is required. Additional
- 14. System Status Condition of key system components Reaction Chamber Pre-Heater Reaction Chamber Cover Reagent Chamber Cover
- 16. An error code appears-Now What ??? Reset Hardware Locate the subsystem in question and perform Diagnostic
- 17. Investigate the Code – OM 11.2 A five digit alphanumeric string having the form axxyz a
- 18. Possible Values for Hardware Errors a = transfer arm r = Reaction Area/Cuvette Wheel d =
- 19. Hardware Error Examples (OM sec 11.3) a0001- Transfer probe failure to find home. Check for object
- 20. System Error Codes P0124 – Empty Sample Cup “ Check Sample at position 1” ID0001- Turned
- 21. System Error Codes S0001- S0002 – Sample Wheel Home and Position errors ( #1 home, #2
- 22. System Error Codes P0124 – Empty Sample Cup “ Check Sample at position 1” ID0001- Turned
- 23. Let’s break that code down! P2304 P=Probe 23=position# 0=R2, 1=R1, 2=Sample, 3=Wash,4=ISE, 5= test/cuvette 4= Bump
- 24. Running Tests after a System Error System beeps to alert you of the error Window appears
- 25. Recovering Cancelled Tests If you end the run, all tests that are in process are cancelled
- 26. Legend of Results Flag codes To view the legend touch the Flag icon on the results
- 29. Measurement Problems They occur during the measurement phase of the test The may affect the resulted
- 30. Measurement Problems- OM 11.11 Chemistries affected Abbreviation Meaning Troubleshooting & Recovery
- 31. Measurement Problems/Flags Appear in the Results Screen on the left side of the Flags column (
- 32. Measurement Flags AS- Air in Sample (ISE”S) Short Sample Clot in sample CD- Calibrator Deviation CV
- 35. Measurement flag- SD (OM Table 11.2) Substrate Depletion - Reagent ran out of substrate for enzymes
- 36. Range Errors User defined pre-set ranges for: QC Ranges - error code = QC Normal Ranges
- 38. Maintenance OM Section 10 Daily Weekly Monthly 6 months As needed EasyRA UI with OM assistance
- 39. Diagnostics EasyRA UI Service software (omit precision dye) ECData
- 40. Diagnostics in Reaction Area Transfer Arm/Probe Fluidics Drawer/ISE’s Reagent Sample Area Note: These diagnostic tests are
- 41. Diagnostics – Reaction area Photometer test The voltage output of the Read and Reference photodiodes The
- 42. Diagnostics – Transfer Arm/Probe Arm Positioning Probe positioning Liquid Level Sensing Let’s do it now!~ OM
- 43. Diagnostics – Fluidics Drawer Dilutor Pump – dilutor pump error or precision test failed/ QC “noisy”
- 44. Diagnostics – Reagent/Sample Area Sample Wheel Barcode Reader RFID Reader Let’s do it! OM pg 11.55
- 45. Precision Test – OM 11.67 The most important test on the An indicator of the performance
- 46. What if my precision test fails?? Check dilutor fittings on the right side of the dilutor
- 47. EasyRA Best Practices Cals Controls Water Reagents Evaporation Environmental E-mail ECData
- 48. Service Training Special Tools Required and supplied by Medica
- 49. Alignment Cuvette Segment Photometer/Liquid level sense/temp cal
- 50. Wash Cup Alignment Tool
- 51. Sample Alignment Tool
- 52. Reagent Wedge Alignment Assembly
- 53. Digital Thermometer
- 54. Two Thermistors
- 55. Foam Insulation Ring
- 56. RFID Test Fixture/Wedge
- 57. Reagent Wedge Base Tool (shim)
- 58. Barcode Test Fixture
- 59. Adjustment Tool for Inductor Liquid level Sensor
- 60. Digital Multimeter with Frequency Counter
- 61. Duct Tape
- 62. Service Software CD
- 63. Service Software Fluidics Drawer – ISE’s Sample/Reagent Area Reaction Area Photometer Mother Board & Daughter Board
- 65. Fluidics Drawer / ISE’s ISE replacement - OM Wash Cup replacement – OM ISE Module replacement
- 66. Dilutor Pump Replacement
- 67. Dilutor Pump Replacement
- 68. Dilutor Pump removal
- 70. Dilutor Pump removal
- 71. Dilutor Pump Replacement
- 72. Dilutor Pump Replacement
- 73. Peri Pump Replacement
- 74. Peri Pump Replacement ISE Module pg29
- 75. ISE Module Replacement
- 76. ISE Module Interface Cable Removal
- 77. ISE Fluidics Drawer
- 78. Sample Reagent Area Sample/Reagent Drive Assembly Fan Assembly (below coolers) RFID Reader ISE Fan Assembly Barcode
- 79. Insulation Cap & Ring
- 80. Sample Reagent Drive Assembly
- 81. S/R Drive Assembly Removal Service Manual
- 82. RFID Reader replacement
- 83. Sample Reagent Area ISE Fan Assembly Barcode Reader Transfer Arm Assembly NOTE! Remove Mid chassis cover
- 84. Removing the Mid Chassis & Parking Cover allows you to……. Remove/replace the ISE fan assembly Remove/replace
- 85. Mid Chassis/Park Cover Removal
- 86. ISE Fan
- 87. Transfer Arm Assembly
- 88. After replacing the Transfer Arm Assembly you must……. Perform an Alignment Perform Liquid Level Sense Calibration
- 89. Reaction Area – Service Manual Reaction Area Drive Assembly Reaction Area Fan/Heater Assembly Reaction Area Thermistor
- 90. After Replacing the Fan and or Thermistor You must…… Perform a Temperature calibration of the Reaction
- 91. Photometer Replacement
- 92. Photometer – Service Manual
- 93. Photometer – Service Manual
- 94. Check Voltages on PCB’s
- 96. Скачать презентацию